In solutions, electric current is carried by moving particles (ions). In solids, electrons are directly transferred from atom to atom. The mobility of the ions is directly proportional to temperature. If the temperature of a solution increases, the conductivity Increases. Therefore, the temperature of the solution must be measured, and a correction factor must be included in the circuitry for temperature changes.
As temperature increases in an aqueous solution, conductivity increases. This will cause the molarity to appear to be higher than it is. This principle is important with regards to the 3081C transmitter because this transmitter will automatically compensate for temperature differences and calculate what the output would be at 25 deg C.
Alternating current is used rather than direct current, with respect to a contact conductivity detector, so that polarization does not occur between the electrodes. The ions are kept mobile between the electrodes while running at typically 1000Hz per cycle. This mobility allows for proper conductivity detection.
The amount of current flow between the electrodes depends on 3 variables. See figure below.
If one was to test the conductivity of distilled water, one would find very little conductivity in the distilled water and could increase the conductivity by adding salt. The level of conductivity will rise if a packet of salt is dumped into the solution being analyzed because salt has electrolytes and electrolytes are directly related to conductivity. Even one single grain of salt will affect the solution. Conductivity is measured in S/cm or Siemens per centimetre. A Siemens is another unit for 1/ohm or a mho.
In order to calibrate a sensor, a solution with a known coductivity, known as standerdization solution, must be used. The conductance of a known standard solution across a 1 cm cube of a material is called specific conductivity.
Potassium Chloride (KCl) is a salt that is easily dissolved in water. It is the most common standardization solution because of its relative harmlessness. Most importantly though, regardless of concentration % By Weight, a known value will be obtained. That is to say that KCl, cannot reach past a known value of µS at 25°C, regardless of the amount added to the solution.
Up to a certain level, the conductivity of a solution increases with concentration. Some solutions such as acids , however, reach a conductivity peak, after which further increase in concentration results in conductivity deceasing.
The electronics need to “see” a resistance between 500Ω to 10,000Ω so that the resistance of the solution is approximately the same as the other arms of the detector circuit. To accomplish this, for fluid with low conductivities the electrodes can be placed close together, or their area can be increased, so the higher conductance (or the lower resistance) can be obtained.
Sampel calcualtion. Given the following data, calculate the slope, correct slope to 25°C and find the K value.
Given
- C=12,856 µS @ 25°C
C=11,167 µS @ 18°C
Calculation
- Therefore the slope = 12,856 µS - 11,167 µS / 25°C -18°C = 241.29 µS/°C
Corrected to 25°C (divide by slope at 25° )= 241.29/12856 = 0.019
K= 100* 0.019 = 1.9%
There is a direct relationship between the cell constant in Transmitter Configuration and conductivity reading. Cell constant acts like a multiplier and directly affects the conductivity reading value.
The concentration effects the conductivity of a solution. Up to certain level (Saturation) the conductivity of a solution increases as the concentration increases. However, some solutions will reach a conductivity peak at a specific concentration after which the conductivity may remain the same or decrease with any further increase in concentration.
The formula for the Cell Constant (Ѳ).
Cell Constant Ѳ = k/ L.
Conductance: The ability of a component to conduct electricity, measured in Siemens.
Resistance: A property of a conductor which the passage of current is opposed, measured in ohms.
Conductance = 1 / Resistance: They are inversely related
A conductivity meter is non-ion selective. It measures all of the ions activity in the solution.
Probe constant is the distance between two electrodes divided by the area of each electrode. Therefore two 1cm² electrodes separated by a distance of 1cm, represents a probe constant of 1. High conductivity solutions require a probe constant greater than 1 and low conductivity solutions require a probe constant of less than 1.
The cell constant if the distance between the electrodes is 0.46 cm, and the surface area is 0.82 cm2?
Cell Constant = L/A = 0.46/0.82 = 0.561
A Toroid (Electrodeless) conductivity sensor is used for corrosive, oily and dirty fluids. A disadvantage is it has lower sensitivity than contact sensors.
Extremely high conductivity requires a sensor with a probe constant grater than 1.0. However, extremely low conductivity requires less than 1.0. The greater distance between the electrodes is defined by the probe constant, the smaller current signal. So we have to use the appropriate probe constant for a specific solution (extremely high conductivity/ extremely low conductivity) in order to provide the most accurate measurement.
Low conductivity solution can be said to have a range of 0.05 to 200 µS/cm and high conductivity solution in the 10 to 20000 µS/cm.
Conductivity Analyzer 1054
Two advantages that using Toroidal Conductivity Measurement has over using Contacting Conductivity Measurement is lower maintenance due to fouling and corrosion resistance and higher range capabilities. The main drawback of Contact measurement is that the sensor is susceptible to coating and corrosion which drastically lowers the conductivity reading. One drawback of Toriodal measurement is that it lacks the sensitivity of a contacting measurement probe and can not be used for low conductivity solutions.
Conductivity is the ability of a solution to conduct an electric current due to the activity of ions in the solution. Ions in a solution are produced from electrolytes (salts, acids, or bases)
The probe constant is a measure of the current response of a sensor to a conductive solution, due to its dimensions and geometry. Its units are cm-1 (length divided by area). The probe constant varies from 0.01 to 50 cm-1. In general the higher the conductivity of a solution the larger the probe constant will have to be.
AC current used instead of DC current because when DC current is used, the ions migrate to their respective polarity nodes whereas with AC current, the ions stay suspended within the liquid.
Contacting conductivity uses a sensor with two metal or graphite electrodes in contact with an electrolyte solution. An AC voltage is applied to the electrodes by the conductivity analyzer and the resulting AC current flowing between the metal or graphite electrodes in a specific solution is measured.
When using the contact probe it is important to remove the air bubble that becomes trapped within the sensor because the air bubble insulates the sensor and will affect the reading.
A Toroidal probe should be selected for operation when the process is dirty or corrosive.
Conductivity is measured with a Toroidal sensor by an AC current induced by the first coil onto the fluid passing through it. As the fluid passes through the second coil it induces a current on that coil. The amount of current induced on the second coil is proportional to the solution conductivity.
Toroidal sensors can be completely coated by a solid or oily contaminant (up to 1cm of thickness) from the process and still not have (significant) effect on the reading. This characteristic can be an advantage depending on the process.For instance, a high turbidity process.
Contacting Conductivity analyzers use a sensor with two metal electrodes in contact with the solution. We apply AC voltage to the electrodes by the conductivity analyzer. As a result, AC current flowing between the electrodes through a volume of solution is used to measure the conductance or the ability to conduct the current.
Weight Percent is when the molecular weight of a solute is unknown or irrelevant, its concentration may be expressed in terms of its weight relative to that of the solution. (Mass of solute / mass of total solution or mixture) x 100
The conductance decreases when a toriod is placed close to a non-conductive surface (like plastic) because the magnetic field travelling from source toroid to measurement toroid is disrupted (see figure below). The lines of flux travel much slower through plastic because the resistance is higher than the fluid, thus creating less inductance in the measurement toroid. This would work opposite in a metal container because metal has less resistance than the fluid.
The three factors that affect conductance are: Velocity of ions in solution, temperature, concentration of ions.